Hot Plate (NINOS) 300? 400 x 300mm

$0.00

  • Model:1-4601-41
Product may vary slightly from image representation.

• Model number: ND-3LA
• Highest temperature (?): 300
• Plate size (mm): 400 x 300
• Temperature settings range: Room temperature +10 - 300?
• Temperature settings: Digital input type
• Temperature distribution width: 4? (*temperature distribution width is indoor environment 25? (no wind, no load, using internal sensor), Reference value when setting to 200?. )
• Temperature control system: Microcomputer type PID control system, ON/OFF control system
• Temperature display system: LCD digital display (setting/measurement temperature simultaneous display, heater output bar graph display)
• Internal temperature sensor: K class thermocouple
• Safety device: Self-diagnosis function (Upper or lower limit temperature abnormality detection, heater abnormality detection, sensor abnormality detection), abnormal heating preventive device (thermostat), overcurrent fuse
• Heater capacity: 1430W
• Additional function: Timer function: Setting range/1 minute - 99 hours 59 minutes (1 minute unit), Program function: Number of steps/maximum 8 steps, repeat times/maximum 9999 times, Setting resolution/temperature 0.1?, time 1 minute, Communication function: PC output (USB), External temperature sensor supported, Upper or lower limit temperature alarm, Alarm signal output, Settings value memory function, Key lock function, Auto-tuning function
• Plate material: Aluminum (ceramic coating)
• Size: 470 x 410 x 101mm
• Power supply: 100VAC 50/60Hz
• Power cord length: Approx 2m
• *Demonstration equipment 1-4601-31, 32, 41, 42 are available.
• Adopts microcomputer PID control method that can regulates temperature with high accuracy.
• Adopts LCD display superior in visibility.
• Integrates 4-types of timer functions(ON?OFF?ON/OFF?SET) as standard. The program function supports up to 8-steps of temperature/time inputs.
• Operation data can be managed on PC easily by using special software for data communication(included in CD-ROM) and communication cable(option)
• Sample temperature can be controlled directly by using external temperature sensor(option).
• Integrates alarm signal output as standard. The signal can be used to stop another equipment or start reserve system upon emergency.
Manufacturers